Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
History

Thu, 22 May 2025 19:45:00 +0000

Type Values Removed Values Added
Description Medtronic MyCareLink Smart 25000 all versions are vulnerable when an attacker who gains auth runs a debug command, which is sent to the reader causing heap overflow in the MCL Smart Reader stack. A heap overflow allows attacker to remotely execute code on the MCL Smart Reader, could lead to control of device. Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
Title Medtronic MyCareLink Smart Heap-based Buffer Overflow
References
Metrics cvssV3_1

{'score': 9.8, 'vector': 'CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H'}

cvssV3_1

{'score': 8.8, 'vector': 'CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H'}


cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published: 2020-12-14T19:18:52

Updated: 2025-05-22T19:37:12.795Z

Reserved: 2020-09-04T00:00:00

Link: CVE-2020-25187

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2020-12-14T20:15:12.653

Modified: 2025-05-22T20:15:21.437

Link: CVE-2020-25187

cve-icon Redhat

No data.